Abstract
Objectives: Evaluation of the precision, accuracy, and user-friendliness of the Technician Immuno I. Calculation of pediatric reference ranges for ferritin and endocrine tests run on Immuno I. Methods: Precision and accuracy were measured using controls and method comparison studies. Pediatric reference ranges were calculated by comparing the Immuno I results for 100 patients with those of the Abbott IMx and TDx. The regression equation obtained was then used to convert the [Mx and TDx reference ranges to reference ranges for the Immuno I. Results: The Immuno I provided both accurate and precise measurement of drugs and endocrine hormones. Pediatric reference ranges were obtained for ferritin and all endocrine tests. Conclusion: The Immuno I is user-friendly and provides reliable measurement of both the drugs tested and endocrine tests on a micro-sample. Reagent and curve stability are excellent.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.