Abstract

We evaluated the MEIA II with blood samples with added tacrolimus (3.0, 5.0, 11.0, and 22.0 microg/L). The assay had acceptable recoveries (99-103%) and intraday imprecision (<16.0%) across the range of concentrations studied, except for the recoveries at 3.0 microg/L (86.3%) and 5.0 microg/L (80.7%). Comparison of liver (n = 116) and renal (n = 113) patient samples measured by MEIA II against HPLC-tandem mass spectrometry (HPLC-MS/MS) found a mean overestimation of 15.6%. From these comparison data it can be calculated that at values of 5 and 20 microLg/L in liver or renal transplant patient samples, measured by HPLC-MS/MS, MEIA II will have the corresponding range estimates of 3.6-7.9 microg/L and 20.9-25.4 microg/L, respectively. No clinically significant difference in results, in terms of overestimation or correlation, was observed between the two transplant groups studied. The MEIA II is an improvement on the previous MEIA I and is suitable for the therapeutic drug monitoring of tacrolimus where HPLC-MS/MS is unavailable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.