Abstract

Background and PurposeTo evaluate the SUV calculation and integration of the gated (4D) PET in the iPlan 4.0 treatment planning software (BrainLAB). Materials and MethodsPhantom and patient data for different tracers were used. Two comparisons were performed for each patient: for the delineated VOI, the maximum value of SUV in iPlan was compared with the results from TrueD software. For 10 patients lesion volumes were defined in both systems for a given SUV threshold and differences were calculated. For four patients examined with respiratory gated PET, SUVmax and volume analysis was performed in each phase of the breathing cycle in the gated and the ungated PET. ResultsMaximum differences of 6% and 10% were found for phantom and patient measurements of SUVmax. For patient data, maximal differences in delineated volume of 10% for ungated and up to 27% for gated PET were found in both systems. ConclusionThis study suggests that for the safe implementation of PET data and delineation algorithms in the radiotherapy planning system, one has to be aware of the differences in SUVs and volumes found in the two systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.