Abstract

Background: A new computer algorithm called Express-View has recently been introduced by Mirocam, but data concerning its application and efficacy are scarce.Objective: To evaluate the lesion detection rate, per-patient sensitivity and the diagnostic accuracy using Express-View.Methods: All patients who performed CE between January 2018 and June 2019, whose indication was obscure gastrointestinal bleeding (OGIB) and with findings on CE, were included. Lesions identified in conventional reading were selected and considered as reference.Results: Eighty-nine patients were included, 50.6% male, with a mean age of 68.4 years-old (±12.3). The Express-View mode detected 85.5% of lesions previously detected by conventional reading (524 out of 613). There were 89 missed lesions, mainly erosions or ulcers (44.9%) and angioectasias (38.2%). The lesion detection rate was found to be lower in the jejunum and ileum compared to extra-small bowel locations and duodenum (p = .04). Although Express-View had a per-patient sensitivity for all lesions of 56.2% and a per-patient sensitivity for all clinically significant lesions of 83.1%, it achieved a diagnostic accuracy of 91%.Conclusions: The per-patient sensitivity for all lesions was shown to be below expectations, although the lesion detection rate, the per-patient sensitivity for all clinically significant lesions, and the diagnostic accuracy were shown to be higher.