Abstract

We have evaluated the technical minimum detection limits of 12 different commercially available kits and the practical minimum detection limits of 11 kits used for the determination of the serum digoxin level by radioimmunoassay. The poor technical ability of many of the kits to distinguish between digoxin levels of 0.1 ng/ml intervals between 0 and 0.5 ng/ml and, additionally, the detection of a digoxin level up to 0.3 ng/ml with some of the kits in glycoside-free sera impose a practical working minimum detection limit of 0.5 ng/ml for the majority of these kits. We, therefore, recommend caution with interpreting or evaluating serum digoxin levels below 0.5 ng/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.