Abstract

This three-center collaborative study was conducted to evaluate samples of Sensititre antimicrobial microdilution panels (GIBCO/INVENEX). Sensititre minimum inhibitory concentrations of 27 bacterial isolates were compared with those obtained by a reference microdilution method. The Sensititre and microdilution minimum inhibitory concentrations were equivalent within +/- 1 dilution in 87.6% of the comparable test results. Intralaboratory reproducibilities of the Sensititre and microdilution endpoints were equivalent with 80.4 and 82.4%, respetively, of on-scale endpoints in absolute agreement. Sensititre was more reproducible among laboratories, with nearly a 10% greater agreement of triplicate results. The Sensititre microdilution test as evaluated gave results which were essentially equivalent to those obtained with a standardized microdilution method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.