Abstract
The main purpose of pre-transcatheter aortic valve implantation (TAVI) cardiac computed tomography (CT) for patients with severe aortic stenosis is aortic annulus measurements. However, motion artifacts present a technical challenge because they can reduce the measurement accuracy of the aortic annulus. Therefore, we applied the recently developed second-generation whole-heart motion correction algorithm (SnapShot Freeze 2.0, SSF2) to pre-TAVI cardiac CT and investigated its clinical utility by stratified analysis of the patient's heart rate during scanning. We found that SSF2 reconstruction significantly reduced aortic annulus motion artifacts and improved the image quality and measurement accuracy compared to standard reconstruction, especially in patients with high heart rate or a 40% R-R interval (systolic phase). SSF2 may contribute to improving the measurement accuracy of the aortic annulus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.