Abstract

The mandible is the most injured part of the facial skeleton, and 25-40% of mandibular fractures involve the condyle process. The aim of this study is to answer the question of the relationship between screw pullout and/or plate fracture during osteosynthesis. We tested polyurethane models of mandibles whose condylar process was cut (simulating a fracture) and fused using plates and screws. A total of 672 plates were tested. A total of 25.6% of them were fractured during the test, with most being fractures of the base of the condyle. More screws (81.97%) are pulled out from the ramus than from the condyle-69.15%. The gold standard in the osteosynthesis of condylar fractures is two straight plates. Other than these, there is no one-size-fits-all plate for every type of fracture. Plates fixed with fewer screws (smaller plates used in higher-lying fractures) are more likely to result in screw pullout. On the other hand, in plates fixed with more screws, plate fracture is more common.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call