Abstract

ObjectiveWe aimed to evaluate the safety and efficacy of radiation dose reduction with a new-generation biplane angiocardiography system in patients undergoing transcatheter isolated patent ductus arteriosus (PDA) closure. Materials and methodsFifty pediatric patients who underwent transcatheter PDA closure were randomly divided into two groups as normal radiation dose and low dose. Patients who required additional procedures other than PDA closure were excluded. PDA closure was performed according to the angiographic measurement of the defect. After the procedure, age, weight, sex, PDA measurements, and radiation measurements such as dose-area product (DAP, Gy.cm2) and air kerma (AK, mGy) were compared between the groups. ResultsThere was no statistically significant difference between the groups in age, sex, weight, PDA diameter, PDA type, device used, and device diameter (p > 0.05). While there was no statistically significant difference between the groups in terms of cine recording, number of recorded images, and fluoroscopy time (p > 0.05), there was a statistically significant difference between the total DAP, cine and fluoroscopy DAP, total AK, frontal and lateral tube AK, and DAP/kg (mGy.m2/kg) measurements (p < 0.05). ConclusionTranscatheter PDA closure with a low radiation dose is as effective as that with a normal radiation dose. The radiation dose received by the patient during the procedure was significantly reduced. With the vision provided by this study, it seems possible to work with a low radiation dose in other groups of patients.