Abstract

In the application of position sensitive radiation detection, the center of gravity method is often used to obtain better spatial resolution. In order to obtain the resolution degradation caused by the readout electronics, an evaluation method is put forward in this paper. The principle of the method is to simulate the signal distribution on sense wires of the MWPC (Multiwire Proportional Chamber) detector. The evaluation software quantizes the simulated data and drives the digital to analog converter to generate test signals. The readout electronics is evaluated with these stimulus signals according to the readout scale. An evaluation board based on this method is implemented. Tests are carried out in the laboratory to evaluate a readout circuit for the MWPC detector. The position resolution contributed by the electronics is better than 15 μm when the swtiching frenquency of the switch array is 12.5 MHz. The test results proved that the evaluation method is correct and practical. It can greatly simplify the evaluation of the readout electronics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.