Abstract
A modified version of the QuEChERS method has been evaluated for the determination of 21 pharmaceuticals and 6 personal care products (PPCPs) in drinking-water sludge samples by employing ultra high liquid chromatography–tandem mass spectrometry (UPLC-MS/MS). The performance of the method was evaluated through linearity, recovery, precision (intra-day), method detection and quantification limits (MDL and MQL) and matrix effect. The calibration curves prepared in acetonitrile and in the matrix extract showed a correlation coefficient ranging from 0.98 to 0.99. MQLs values were on the ngg−1 order of magnitude for most compounds. Recoveries between 50% and 93% were reached with RSDs lower than 10% for most compounds. Matrix effect was almost absent with values lower than 16% for 93% of the compounds. By coupling a quick and simple extraction called QuEChERS with the UPLC-MS/MS analysis, a method that is both selective and sensitive was obtained. This methodology was successfully applied to real samples and caffeine and benzophenone-3 were detected in ngg−1 levels.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.