Abstract

We present an experiment in which four different FG5 absolute gravimeters (AG) were operatedsimultaneously alongside a superconducting relative gravimeter (SG). We demonstrate that 0.1 % precision can be achieved on the calibration factor of the SG by comparison with AG measurements, independently of the FG5 instrument used for the calibration and of the offsets among the FG5 absolute values. This experiment demonstrates the robustness of using any FG5 absolute gravimeter to calibrate any SG. This result is of value to geoscientists analysing data from (a) globally distributed SGs which most probably have been calibrated using different instruments; or (b) any individual SG calibrated with different FG5 absolute gravimeters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.