Abstract
We evaluated the performance of the PG-Numeric (PGN; Isolab, Inc.) enzymatic assay for phosphatidylglycerol (PG) adapted for semi-automated analysis on the Cobas-Bio (Roche) centrifugal analyzer. We evaluated precision, linearity, and potential interference from endogenous glycerol. In addition, we compared the results of the PGN assay with those for two other PG assays. Amniostat-FLM (AFLM; Hana Biologics) and phosphatidylglycerol/sphingomyelin ratio (PG/S), on amniotic fluid samples from 51 patients. Within-run and total CVs for the PGN test (n = 80) were respectively 8.6% and 11.4% for PG at 2.6 mumol/L, and 4.5% and 6.2% at 7.3 mumol/L. The PGN test exhibited no significant deviation from linearity between PG concentrations of 0.4 and 15 mumol/L. Endogenous glycerol up to 500 mumol/L did not affect PGN results significantly. Results of the PGN assay compared favorably with those of the other PG assays for tests of fetal lung maturation. The Cobas-Bio adaptation of the PGN test performed well analytically and clinically, and allowed the determination of PG in less than 1 h.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.