Abstract

Deep Brain Stimulation (DBS) is an established therapy for many movement disorders. DBS entails electrical stimulation of precise brain structures using permanently implanted electrodes. Following implantation, locating the electrodes relative to the target brain structure assists patient outcome optimization. Here we evaluated an open-source automatic algorithm (PaCER) to localize individual electrodes on Computed Tomography imaging (co-registered to Magnetic Resonance Imaging). In a dataset of 111 participants, we found a modified version of the algorithm matched manual-markups with median error less than 0.191 mm (interquartile range 0.698 mm). Given the error is less than the voxel resolution (1 mm3) of the images, we conclude that the automatic algorithm is suitable for DBS electrode localizations.Clinical Relevance- Automated DBS electrode localization identifies the closest electrode to the target brain structure; allowing the neurologist to direct electrical stimulation to maximize patient outcomes. Further, if none of the electrodes are deemed suitable, localization will guide re-implantation.