Abstract

Cyclosporine (CyA) has a narrow therapeutic index. Determination of CyA concentrations correlate with rejection or adverse effects like nephropathy. Cyclosporine is assayed based on either chromatographic or many different immunoenzymologic techniques. The investigators evaluated a new heterogeneous immunoassay of CyA on RxL Dimension®. The pretreatment step is automatically performed in the apparatus. Linearity, intra- and interday precision, limit of quantification, dilutions, and stability of the equipment were compared with the EMIT method for patient determinations. The heterogeneous immunoassay showed a good linearity between 0 and 500 ng/mL, and intra- and inter-day precision with a coefficient of variation below 9.2%. The investigators observed reproducible and accurate dilutions of high concentrations (500 to 2000 ng/mL). The correlation with the EMIT technique was valid: ACMIA = 0.964 EMIT + 0.156 ( r = .96) for different types of transplant (n = 116). Finally, this new system improves the determination of CyA concentrations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.