Abstract

The direct inoculation of the Minitek anaerobe identification system (BBL Microbiology Systems, Cockeysville, Md.) from positive blood cultures was compared with subculture and Minitek results obtained using the manufacturer's recommended procedures. A total of 40 clinical anaerobic blood cultures were processed for rapid identification utilizing bacterial pellets obtained by centrifugation. Of these cultures, 30 yielded pure isolates of anaerobic rods that were used for comparison. In 87% of the pure cultures, identification from the direct inoculum method was identical to the routine procedure using Minitek biochemicals. When the additional test for lecithinase and lipase production was included, the identification agreement was 97%. Direct identification of anaerobic rods from blood cultures utilizing the Minitek system is reliable, easy to perform, and can provide a complete identification in 24 h from the time a blood culture showed growth.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.