Abstract
The aim of the study was experimentally to evaluate the capability and reliability of laser Doppler flowmetry (LDF) in conditions of circulatory deficiency, by correlating it to flow-related parameters measured by a radionuclide-imaging technique and using 99m-Tc red blood cells (RBCs). For this purpose, a pedicle island flap in the sheep was used, with well-perfused proximal parts and with evident stasis in the distal third of the flap. No correlation was found between results obtained with the two techniques. In regions with evident stasis, falsely high LDF readings were recorded. This may be due to a back-and-forth motion of the RBCs under the probe, rather than to true flow. It was concluded that, while LDF seems reliable in detecting complete arterial occlusion, it is unreliable in predicting either complete venous occlusion or partial obstruction of the flow to and from the flap. Clinical use for this purpose cannot be recommended.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.