[purpose] Using the liquid chromatography-Ultraviolet method ,the monosomic Ginsenosides Rb1 content is detected, and evaluated the measuring uncertainty of the monosomic Ginsenosides Rb1 content. [method] Analyzing the main source of the uncertainty components, finally evaluating and calculating the uncertainty components. [result]The results shows that, when the monosomic Ginsenosides Rb1 content is in the range of 0.02-0.12mg/ml, there is a good linear relation, and the linear equation is A=13651.52c+0.1430, r=0.9999 (coefficient correlation) For example, when the content of monosomic Ginsenosides is 0.3421%, the expanded uncertainty of monosomic Ginsenosides measurement is U=0.0236%(K=2). [conclusion]This study provides a reliable theoretical basis for the effective controlling the quality of monosomic Ginsenosides Rb1 content measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.