Abstract

In this evaluation a mastascanelite image analysis system is compared with manual measurement of disc diffusion inhibition zones. Data for 213 diverse organisms and a total of 1679 organism/antibiotic combinations gave an overall correlation coefficient of 0.988. The mean difference between readings was +0.425 mm, with 95% confidence limits of ±2.94 mm, and the majority of scanned zones (97.51%) fell within ±3 mm of the manual measurement. The mastascanelite system forms part of a laboratory suite and was found to be objective, accurate and rapid, reading and interpreting each plate in less than a second. Interfacing to the laboratory computer system facilitated data handling and performance control.