Abstract

The present paper is a report of a project to develop an automated auditory screening test for infants six to 18 months of age. The first year of the project was devoted to developing equipment and test procedures; the second year was concerned with testing the effectiveness of the equipment and procedures on an actual population of six to 18 month old infants. Two-hundred and fifty infants were screened auditorily as part of a county health department child development clinic. The pass/fail results of the screening test were evaluated in terms of physical and developmental examination following the screening and by means of a case review of the child's previous history. The results indicate that the procedure under investigation can be used to differentiate the normal hearing infant from the infant with possible hearing problems. It is shown by the test environment in which this study was conducted that the procedure reported can be successfully incorporated into a public health program, i.e., child development clinics or EPSDT programs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.