Abstract

Antinuclear antibodies (ANA) are widely detected by immunofluorescence on HEp-2 cells in patients with connective tissue diseases and other pathological conditions. We evaluated the first-automated chemiluminescence immunoassay for the detection of ANA (LIAISON ANA screen, DiaSorin). This study was carried out simultaneously in two laboratories by testing 327 patient samples with clinically defined connective diseases, 273 routine samples for ANA screening, and 300 blood donors. A total of 268 out of 337 IIF-positive sera were positive with LIAISON ANA screen (79.5% of agreement) and 240 out of 263 IIF-negative sera were negative with LIAISON ANA screen (91.2% of agreement). After resolution of discrepant results, the concordance reached, respectively, 94.9% and 98.8%. The specificity was 99.3% and the sensitivity was 94%. Unlike results obtained by other ANA screening assays, we observed acceptable sensitivity and specificity. Despite the presence of HEp-2 cell extract, we failed to detect some antibodies as antinucleolar, antinuclear envelope, and antiproliferating cell nuclear antigen. This automated assay allows quick process to results and exhibits satisfactory sensitivity for the detection of the main ANA specificities of connective tissue diseases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.