Abstract

The kinetic parameters of the first, third and fifth peaks of CaF2:Tm (TLD-300) dosemeters were determined experimentally using methods based on initial rise, peak shape, different heating rates, isothermal decay and numerical fitting method and a comparison was made between the results obtained. The average activation energy values determined by the methods used here were 0.72, 1.16 and 1.77 eV, with mean values for the frequency factor of 2.58*108, 2.89*1013 and 6.16*1016 s-1 for the first, third and fifth peaks respectively. The values of activation energy obtained for each peak using the different methods covered a spread of about 4%. However, the spread for each of the frequency factors was very high.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.