Spinal instrumentation using transpedicular screws has been used for decades to stabilize the spine. In October 2018, an intraoperative CT system was acquired in the Neurosurgery service of the University Hospital Complex of Vigo, this being the first model of these characteristics in the Spanish Public Health System, so we began a study from January 2015 to December 2019 to assess the precision of the transpedicular screws implanted with this system compared with a control group performed with the classical technique and final fluoroscopic control. The study was carried out in patients who required transpedicular instrumentation surgery, in total 655 screws were placed, 339 using the free-hand technique (Group A) and 316 assisted with intraoperative CT navigation (Group B) (p>0.05). Demographic characteristics, related to surgery and the screw implantation grades were assessed using the Gertzbein-Robbins classification. 92 patients were evaluated, between 12 and 86 years (average: 57.1 years). 161 thoracic screws (24.6%) and 494 lumbo-sacral screws (75.4%) were implanted. Of the thoracic screws, 33 produced a pedicle rupture. For the lumbo-sacral screws, 71 have had pedicle violation. The overall correct positioning rate for the free-hand group was 72.6% and for the CT group it was 96.5% (p<0.05). The accuracy rate is higher in thoracic-lumbar instrumentation in the navigation group versus free-hand group with fluoroscopic control.