Abstract

Two hundred and eighty Rapid Plasma Reagin (RPR) positive sera with an emphasis on cases with negative and borderline positive Treponema pallidum haemagglutination assay (TPHA) results were selected. Modified TPHA (M-TPHA) and fluorescent treponemal antibody absorption (FTA-abs) tests were used for comparison. One hundred and twenty five samples were TPHA negative, of which 78 and 69 cases were also negative by M-TPHA and FTA-abs, respectively. Eighty one sera negative by TPHA at a titre of 1/80 and positive at 1/40, considered to be negative according to the manufacturer's instructions, were also negative by M-TPHA (n = 11) and by FTA-abs (n = 1). Fifty borderline positive TPHA specimens gave one negative result by both M-TPHA and FTA-abs. The remaining 24 sera were positive by all three tests. Because of the high percentage of TPHA negative results among the positive RPR sera which became reactive when rechecked by the FTA-abs, it is concluded that as a confirmatory test the TPHA should be used not instead of but in addition to the FTA-abs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.