Abstract
A cross-sectional study was conducted to evaluate the effectiveness of manual and automated dialyzer reprocessing. Dialyzers were filled with fluid thioglycollate medium from blood and dialysate chambers after being reprocessed and chemically sterilized with 0.2% peracetic acid. They were incubated for 14 days at 35°C ± 2°C, and microbiologic analysis was performed. Microorganisms were identified in 3 of the 11 samples (27.3%) from the blood chambers: Sphingomonas paucimobilis (2/3) and Penicillium spp (1/3) and in 11 of the 11 samples (100%) from the dialysate chambers: S paucimobilis (7/11), Stenotrophomonas maltophilia (4/11), Pseudomonas aeruginosa (3/11), Candida spp (1/11), and Acinetobacter baumannii (1/11). Of the 4 manually reprocessed dialyzers, gram-positive bacillus were identified in 1 sample (25%) from the blood chamber, and Bacillus spp and Burkholderia spp were identified in 1 sample (25%) from the dialysate chamber. The dialyzers reprocessing can pose risks safety because of exposure patient to microorganisms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.