Abstract
A sensitometric study of Kodak XV and EDR-2 radiographic films (Eastman Kodak Company, Rochester, NY) was performed using photons ranging from 75 kV to 18 MV and electrons ranging from 6 to 20 MeV. To investigate the applicability of the EDR-2 film for clinical radiation dosimetry, percentage depth-doses, profiles and distributions in open and dynamically wedged fields were measured using film and compared with data from a linear diode. Moreover, conventional quality assurance dose parameters were measured, including open-field dose profiles to determine flatness and symmetry of photon and electron beams. Finally, film was employed to validate dose distributions produced by complex computerised treatment planning techniques. Our conclusion is that the EDR-2 film is an effective tool for relative dosimetry of photon and electron beams.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.