Abstract

The Early Access STR Kit v1 is designed to detect 25-plex loci with next generation sequencing (NGS) technology on the Ion Torrent PGM™ platform, including 16 of 20 expanded Combined DNA Index System (CODIS) core loci (CSF1PO, D1S1656, D2S1338, D2S441, D3S1358, D5S818, D7S820, D8S1179, D10S1248, D13S317, D16S539, D19S433, D21S11, TH01, TPOX and vWA), 8 non-CODIS core loci (D1S1677, D2S1776, D4S2408, D5S2500.AC008791, D6S1043, D6S474, D9S2157 and D14S1434) and Amelogenin. In this study, we compared the Early Access STR Kit v1 with the Ion Torrent™ HID STR 10-plex to find out its improvements and explored an appropriate analytical threshold to enhance the performance. In addition, seven experiments were conducted to evaluate the Early Access STR Kit v1 such as studies of repeatability, concordance, sensitivity, mixtures, degraded samples, case-type samples and pedigrees. Other than a little discordance (0.95%) with CE-STR results observed at D21S11, NGS-STR results correctly reflected the sample being tested. Repeatable results were obtained from both initial PCRs and emPCRs aside from a few variations of allele coverage. Full profiles could be obtained from 100pg input DNA and >48.84% profiles from 10pg input DNA. Mixtures were easily detected at 9:1 and 1:9 ratios. This system could be adapted to case-type samples and degraded samples. As a whole, the Early Access STR Kit v1 is a robust, reliable and reproducible assay for NGS-STR typing and a potential tool for human identification.