Abstract

Measurement of HbA 1c is key in monitoring diabetic patients in both laboratories and clinical units, where HbA 1c results are used as part of patient education. We have evaluated the DCA Vantage, a new device for immunological assay of HbA 1c. HbA 1c results obtained were evaluated in terms of precision, linearity, specificity and practicability, and were compared with results obtained by a Variant II HPLC method. The method exhibited intra- and inter-assay coefficients of variation lower than 2.6% and 4.0%, respectively, and good correlation with the comparison HPLC method (r2=0.9776). No interference was noted in the presence of labile HbA 1c or carbamylated hemoglobin. The new device exhibited improved practicability characteristics and allowed better sample identification, better management of quality control routines and greater connectivity possibilities compared to the previous DCA 2000 analyzer. This new analyzer exhibited analytical and practical characteristics very suitable for HbA 1c assay for laboratory or point-of-care use according to good laboratory practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.