Abstract
We evaluated the COBAS AMPLICOR polymerase chain reaction (PCR) based test for the detection of Mycobacterium tuberculosis complex in 866 respiratory and non-respiratory samples. Acid-fast staining and culture on Lowenstein-Jensen medium were also performed on all samples. Of the 866 samples tested, 87 (10.0%) were PCR-positive compared to 94 (10.9%) culture positive. There were no false positive results but 7 PCR-negative, culture-positive samples were, considered false negatives after reviewing medical records of patients. A PCR inhibitory rate of 2.0% (17/866) was observed in respiratory samples only. Sensitivity, specificity, and positive and negative predictive values for this test were 92.5%, 100%, 100% and 99.1% respectively. This test is a valuable diagnostic tool for today's mycobacteriology laboratory.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.