Abstract

The COBAS TaqMan HIV-1 test (Roche Diagnostics) was compared with the LCx HIV RNA quantitative assay (Abbott Laboratories), the Versant HIV-1 RNA 3.0 (bDNA) assay (Bayer) and the COBAS Amplicor HIV-1 Monitor v1.5 test (Roche Diagnostics), using plasma samples of various viral load levels from HIV-1-infected individuals. In the comparison of TaqMan with LCx, TaqMan identified as positive 77.5% of the 240 samples versus 72.1% identified by LCx assay, while their overall agreement was 94.6% and the quantitative results of samples that were positive by both methods were strongly correlated ( r = 0.91). Similarly, in the comparison of TaqMan with bDNA 3.0, both methods identified 76.3% of the 177 samples as positive, while their overall agreement was 95.5% and the quantitative results of samples that were positive by both methods were strongly correlated ( r = 0.95). Finally, in the comparison of TaqMan with Monitor v1.5, TaqMan identified 79.5% of the 156 samples as positive versus 80.1% identified by Monitor v1.5, while their overall agreement was 95.5% and the quantitative results of samples that were positive by both methods were strongly correlated ( r = 0.96). In conclusion, the new COBAS TaqMan HIV-1 test showed excellent agreement with other widely used commercially available tests for the quantitation of HIV-1 viral load.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.