Abstract

ABSTRACTThe Breathalyzer® 7410-CDN with Printer was evaluated according to the CSFS Alcohol Test Committee standards (1995) for approved instruments. The Breathalyzer® 7410 is physically similar to the Alcotest® 7410 Approved Screening Device and, similarly, uses an electrochemical sensor for measuring alcohol. The Breathalyzer® 7410, however, is programmed to function as an evidential breath alcohol analyzer and is accompanied by a printer. Data from the Breathalyzer® 7410 is optically transmitted to the printer. The instrument did not respond to acetone vapours, but did respond to other alcohols such as methanol and isopropanol. The instrument showed good precision and accuracy when tested with simulated breath alcohol vapours in the range of 50 to 350 milligrams of alcohol in 100 millilitres of solution (mg%). Human subject testing (n = 11) showed an excellent correlation with an Approved Instrument (Intoxilyzer® 5000C). Blood/breath correlations with six of the subjects showed that the Breathalyzer® 7410 underestimated the corresponding blood alcohol concentration in a range of 10 to 30 miligrams in 100 millilitres of blood.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.