Abstract
The sensitivity and specificity of the Bio-EnzaBead test for syphilis and the fluorescent treponemal antibody-absorption (FTA-ABS) test were determined by examining 262 serum samples, including 202 serum samples from patients with confirmed syphilis in various stages. Overall correlation with patient history was 95.8% with both tests. False-negative Bio-EnzaBead tests occurred in 9 of 86 (10.5%) cases of late-latent syphilis (greater than 2 years) and in 1 of 38 (2.6%) cases for which the stage of disease could not be determined. False-negative FTA-ABS tests occurred in 5 of 86 (5.8%) cases of late-latent syphilis (greater than 2 years) and in 2 of 38 (5.3%) cases for which the stage of disease could not be determined. One false-positive test occurred with Bio-EnzaBead, and the cause could not be determined. The reproducibility of the Bio-EnzaBead test was excellent when spectrophotometric readings were calibrated against either air or substrate blanks. The Bio-EnzaBead test for syphilis is a suitable alternative to the FTA-ABS test.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.