Abstract

Aim: To evaluate the performance of an automated BACTEC MGIT 960, a non-radioactive, non-invasive liquid culture system for cultivation of M. tuberculosis complex in terms of recovery rate and time. Materials and Methods: From March 2005 to December 2007, 14,597 specimens were processed using the MGIT 960 system and the results were compared with conventional L.J medium. We standardised ρ-nitro benzoic acid (PNBA) assay on MGIT 960 TB system for identification of M. tuberculosis complex and evaluated its usefulness by comparing the results with an in-house molecular assay and sequencing. Results and Discussion: Of the total 6143 (42%) isolates positive for M. tuberculosis complex, 6015 (41%) were positive by MGIT 960 TB system. In contrast, 3526 (24%) M. tuberculosis complex isolates grew on the conventional L.J medium. The mean turn around time for mycobacterial growth in smear-positive specimens was nine days for MGIT 960, and 38 days for L.J. medium whereas in smear negative specimens it was 16 days by MGIT vs. 48 days by L.J. Conclusion: MGIT 960 system with PNBA assay for identification of M. tuberculosis complex is a rapid and useful method in laboratories processing a large number of specimens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.