Abstract

HIV-1 RNA quantitation in plasma, or virus load testing, is the primary method by which the response to antiretroviral therapy is monitored. Here we describe evaluation of the Aptima HIV-1 Quant Dx assay (Aptima) performed on the automated Panther system. The clinical performance of Aptima was compared to that of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test v2.0 (CAP/CTM) using 162 EDTA plasma samples collected from patients undergoing HIV-1 monitoring. Overall agreement was 84.0% (136/162), with a kappa statistic of 0.723 (standard error, 0.047; 95% confidence interval [CI], 0.630 to 0.815), indicating substantial agreement. Using the 86 clinical samples quantifiable by both methods, Passing-Bablok regression revealed a regression line of Y = (1.069 × X) - 0.346 (95% CI of the slope [1.003 to 1.139] and intercept [-0.666 to -0.074]), and Bland-Altman analysis demonstrated a mean difference (Aptima-CAP/CTM) of -0.075 log10 copies/ml (95% limits of agreement of -0.624 to 0.475), consistent with negative bias. Comparison of Aptima results for paired dried blood spot (DBS) and plasma specimens archived from participants in the Peninsula AIDS Research Cohort Study (PARC) demonstrated an overall agreement of 94.7% (90/95) when 1,000 copies/ml was used as the threshold. In conclusion, the Aptima HIV-1 Quant Dx assay provides a suitable alternative for HIV-1 monitoring in plasma and DBS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.