Abstract

Objective To evaluate the analytical performance of the Nova StatSensor creatinine meter. Design and methods Imprecision, linearity, patient correlation and interference studies were completed. Results Total imprecision was 4.5–9.1%CV and a linear measurement range of 93–863 µmol/L (1.05–9.76 mg/dL) was verified. Whole blood creatinine results correlated well with laboratory plasma measurements ( R 2 = 0.9328) but exhibited a negative proportional bias. In vitro, high levels of creatine and urea falsely elevated creatinine measurement. Conclusions The creatinine meter provides reliable measurement across a clinically relevant range and has potential use in point-of-care testing.