Abstract

Dry reagent technology and reflectance photometry are combined in the Ames Seralyzer to offer a solid-phase plastic strip assay methodology. Light reflected from serum placed on the antibody-impregnated strip is quantitated with a two-point standard curve to display a theophylline result in 90 seconds. The accuracy and precision of the Seralyzer for serum theophylline are compared to the enzyme-multiplied immunoassay technique (Emit) which is commonly used to determine drug concentrations. Liquid calibrators (5 and 25 micrograms/ml) and liquid-spiked sera (10, 15, and 20 micrograms/ml) were measured six times daily simultaneously by both methods for twenty days. Eighty patient samples (theophylline range: 0-33 micrograms/ml) were measured twice each during the twenty days. Acceptable assay limits were established and maintained by measuring 15 micrograms/ml spiked sera prior to and during the evaluation period. All within-run and between-run means for the Seralyzer and Emit were within +/- 1 microgram/ml of the spiked value. All ranges of within-run and between-run means were within +/- 2 micrograms/ml of the spiked value. Mean between-run coefficients of variation for the Seralyzer at the 5, 10, 15, 20, and 25 micrograms/ml values were 11.5, 6.7, 4.9, 4.6, and 4.9 percent, respectively. Linear regression on the 80 patient samples gave a slope of 1.02, intercept of -0.28, and correlation of 0.984. The Seralyzer's accuracy and precision compare favorably with Emit for theophylline determinations greater than 7.5 micrograms/ml.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call