Abstract
DNA typing of short tandem repeat (STR) loci with automated real-time analysis of the fluorescent polymerase chain reaction (PCR) products has given forensic DNA analysis a new dimension. In the present work, the ALF DNA sequencer was evaluated for automated size determination of tetra-nucleotide STRs at high speed. Short gel plates, with a well-to-laser distance of 10 cm, allowed for the analysis of four STR loci (HUMvWF, HUMTH01, D21S11 and HPRT) in one gel lane in less than 75 min. Allele size determination was done with two external allelic ladders for each locus. Lane-to-lane variations were overcome by the inclusion in each lane of two fluorescent PCR products of constant size (123 and 375 bp) that migrated below and above the multiplex of the four STR loci. The accuracy of sizing and allele detection within and between different gels was high (99.89%) for all four STR systems investigated and the gels could be reloaded without a decrease in accuracy of the allele size estimation. This way, the throughput of the system was increased, which is of interest for linkage studies, gene mapping, and population diversity studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.