Abstract

The purpose of this study was to prospectively test the accuracy of computer-aided orthognathic surgery comparing the virtual surgical planning with the three-dimensional (3D) outcome. Patients that underwent computer-assisted orthognathic surgery were retrospectively evaluated. The postoperative results were compared with the surgical plan, superimposing the postoperative computed tomography (CT) scan onto the virtual plan. Surface-based superimpositioning of the postoperative CT scan onto the 3D preoperative plan was carried out to visualize the discrepancy between preoperative virtual plan and postoperative 3D CT result. A total of 17 consecutive patients that underwent two-jaw computer-assisted orthognathic surgery were enrolled in the study.The average linear differences for selected points were <1 mm in 12 patients out of 17. In 5 patients out of 17, the average differences for selected points were <2 mm. An overall high degree of accuracy between the virtual plan and the postoperative result was found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.