Abstract
Three-dimensional (3D) quantification using an interleaved Look-Locker acquisition sequence with a T2 preparation pulse (QALAS) is a quantitative sequence used to measure relaxation times. The accuracy of the relaxation time measurement of 3D-QALAS at 3.0 T and the bias of 3D-QALAS have not yet been assessed. The purpose of this study was to clarify the accuracy of the relaxation time measurements using 3D-QALAS at 3.0 T MRI. The accuracy of the T1 and T2 values for 3D-QALAS was evaluated using a phantom. Subsequently, the T1 and T2 values and proton density of the brain parenchyma in healthy subjects were measured using 3D-QALAS and compared with those of 2D multi-dynamic multi-echo (MDME). In the phantom study, the average T1 value of 3D-QALAS was 8.3% prolonged than that for conventional inversion recovery spin-echo; the average T2 value for 3D-QALAS was 18.4% shorter than that for multi-echo spin-echo. The in vivo assessment showed that the mean T1 and T2 values and PD for 3D-QALAS were prolonged by 5.3%, shortened by 9.6%, and increased by 7.0%, respectively, compared with those for 2D-MDME. Although 3D-QALAS at 3.0 T has high accuracy T1 value, which is less than 1000 ms, the T1 value could be overestimated for tissues with it longer than that T1 value. The T2 value for 3D-QALAS could be underestimated for tissues with T2 values, and this tendency increases with longer T2 values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.