Abstract

Accurate orientation of acetabular and femoral components are important during THA. However, no study has assessed the use of the CT-based fluoro-matched navigation system during THA. Therefore, we have evaluated the accuracy of stem orientation by CT-based fluoro-matched navigation. The accuracy of stem orientation by CT-based fluoro-matched navigation was assessed by postoperative CT data. Furthermore, we compared the postoperative stem orientation with the intraoperative registration errors. The average antetorsion error of the stem (navigation records - postoperative CT) was -0.5° ± 5.2°. The stem valgus error was 0.4° ± 2.7°. The accuracy of the navigation record for the orientation of the stem valgus was dependent on the intraoperative registration errors. The clinical accuracy of CT-based fluoro-matched navigation is adequate for stem alignment orientation, and the intraoperative verification of registration errors is valuable for checking the accuracy of stem orientation by navigation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.