Abstract

Development of the ARCHITECT Toxo IgM assay has been done to assist the clinician in acute Toxoplasma gondii infection detection, especially in pregnant women. Its use, in conjunction with ARCHITECT Toxo IgG and Toxo Avidity assays, will provide an array of assays particularly useful in the monitoring of pregnant females to determine the risk of maternal transmission of the parasite. Specificity results from 2 testing sites, using populations of pregnant females, hospital patients, and blood donors, demonstrated that the assay has an overall resolved relative specificity of 99.89% (confidence interval, 99.68–99.98%). Relative specificity for pregnant female specimens was 99.95% ( n = 2031). Excellent seroconversion sensitivity was observed for the ARCHITECT Toxo IgM assay, which was similar to the Abbott AxSYM Toxo IgM assay (Abbott Laboratories, Abbott Park, IL). In more than 90% of the panels tested, the 1st bleed detected in the serial bleeds was the same for both assays.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.