Abstract

Myoglobin, CK-MB, and Troponin I (cTnI) are cardiac muscle necrosis markers that are useful for detecting acute myocardial infarction (AMI). The Stratus CS (Dade Behring, Inc.) is a discrete fluorimetric immunoassay analyser designed for the determination of the three cardiac markers from a single sample of whole blood or plasma. Overall analytical performances of the Stratus CS provided by Dade Behring were evaluated according to the French Society of Clinical Biology guidelines. Within-run imprecision (n = 20) for the three parameters at three levels gave values under 5%, whereas CVs for between-run imprecision (n = 20) were under 6%. The sensitivities were 0.03 microg/L for cTnI and 0.4 microg/L for CK-MB. Linearities extended from 0-50 microg/L for cTnI, 0-140 microg/L for CK-MB, and 1-900 microg/L for myoglobin. The results, particularly those obtained on whole-blood samples, correlated well with those obtained on Stratus II. We did not find any interference with haemolysis, icterus, or lipemia. The system was very easy to use, and fulfills the requirements for the analysis of the three cardiac markers in patients with acute chest pain in emergency situations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.