Abstract

This paper presents the effects of conductive medical implant on energy absorbed by the human body and the testicular area when exposed to near field electromagnetic radiation. A dipole antenna is used as the radiating source and it is placed in front of the trousers pocket. Two types of medical implants are used in this study: intramedullary nail and bone plate. Numerical simulations are performed by means of CST Microwave Studio. Results are discussed in terms of changes in SAR values due to the presence of conductive medical implant at 0.4, 0.9, 1.8 and 2.4 GHz. The results have indicated that the conductive intramedullary nail that is located inside the femur significantly increases the SAR. Maximum enhancement in SAR is found when the length of the intramedullary nail is approximately one wavelength of the respective frequency tested. The measurement results indicate good agreements with the simulation results at 2.4 GHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.