Abstract

An electrochemical biosensor for H2O2 detection was developed by using soybean peroxidase-copper phosphate mediated organic inorganic hybrid (OIH). The characterization of OIH was carried out by FESEM and FTIR. FESEM analysis showed a flower-like porous morphology of the OIH and FTIR confirmed the presence of soybean peroxidase and copper phosphate in the OIH. For sensor development, the paste of OIH was formulated in the presence of phosphate buffered saline (PBS) and was screen printed on indium tin oxide (ITO) coated glass substrate. Cyclic voltammetry analysis showed that the developed biosensor could detect H2O2 in the linear range of 20–100 ​μM with R2 value of 0.963. The limit of detection (LOD) and sensitivity values calculated for H2O2 are 0.19 ​μM and 27.44 μA/(μM.cm2) respectively. Along with cyclic voltammetry experiments, electrochemical impedance spectroscopy (EIS) analysis was also carried out to study the sensing mechanism. The developed biosensor showed better selectivity towards H2O2 when tested against d-glucose, l-tyrosine, l-lysine, and l-ascorbic acid.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.