Abstract

The aim of our study was to evaluate whether software-based artifact reduction can achieve an improved image quality, using intraoperative 3D imaging in spinal surgery. A total of 49 intraoperative 3D image datasets of patients, who underwent surgery with pedicle screw placement, were retrospectively evaluated. The visibility of anatomical structures and the diameter of the pedicle screws were examined, with and without the application of the artifact reduction software. All software prototypes can improve the visibility of anatomical structures (P < 0.01), except MAR (metal artifact reduction) combined with IRIS (iterative reconstruction in image space) (P = 0.04). The algorithms MAR and MAR-2 can reduce the blooming artifacts significantly (P < 0.01), but SL (Shepp & Logan) cannot (P = 0.08-0.988). In summary, software-based artifact reduction for intraoperative 3D datasets can improve the current image quality. Additional information regarding the implant placement and the fracture reduction is therefore generated for the surgeon.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.