Abstract

Quantitative studies were conducted to evaluate the efficiency of the slit sampler in collecting airborne Serratia marcescens and Bacillus subtilis var. niger, and to compare it with the collecting efficiency of the all-glass impinger AGI-30. The slit sampler was approximately 50% less efficient than the AGI-30. This ratio remained the same whether liquid or dry cultures were disseminated when the sample was taken at 2 min of aerosol cloud life. At 30 min of aerosol cloud life, this ratio was approximately 30% for B. subtilis var. niger. S. marcescens recoveries by the slit sampler were, however, only 17% lower than the AGI-30 at 30 min of cloud age, indicating a possible interaction involving the more labile vegetative cells, aerosol age, and method of collection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.