Abstract

Abstract Silicone tubing is frequently used for gas exchange in cell culture systems, due to its biocompatibility and high permeability to CO2 and O2. In cell culture chambers, medium pH and oxygen levels are often maintained by gas exchange through a coil of silicone tubing. Culture medium is recirculated between the gas exchanger and the culture chamber which contains a suspension of cells. We report that the type of agent used for silicone vulcanization (peroxide or platinum) can markedly affect its biocompatibility, and that tobacco cell culture represents a particularly sensitive indicator of tubing cytotoxicity. Under the conditions studied (cell suspension maintained with forward–reverse flow and stirring), peroxide-cured silicone tubing was toxic to the tobacco BY2 cell culture, in contrast to the platinum-cured silicone tubing that was completely biocompatible. Upon further investigation by mass spectrometry, it was determined that a component with a molecular mass of 288 Da, possibly a tetrachlo...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.