Abstract

In order to provide information on the suitability of commonly used gas storage vials for air transport, we tested two vial types on their ability to preserve defined nitrous oxide concentrations and excess pressure when exposed to low pressure, low temperature and puncture by needles. Unlike in Crimp Cap vials, in Exetainers no nitrous oxide loss following low pressure storage was detectable. Tightness of Exetainers following multiple puncture was best using a small needle diameter. Pressure loss following 5, 10, or 25 punctures was lowest in the Exetainers. We conclude that Exetainers are suitable for storing gas samples for an extended period of time during aircraft transport.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.