Abstract

ABSTRACTScreening of blood for human T-cell lymphotropic virus type 1 and type 2 (HTLV-1 and -2, respectively) is important to diagnose and prevent infection and ensure the safety of blood supplies. The Elecsys HTLV-I/II assay is a newly developed, electrochemiluminescence screening assay for the detection of HTLV-1/2 infection. The sensitivity and specificity of the Elecsys HTLV-I/II assay were determined using well-characterized HTLV-1/2-positive serum and plasma samples and routine diagnostic and blood donor samples expected to be HTLV negative, respectively. These results were compared with those for at least one of the following CE-marked assays at seven independent laboratories and the Roche Diagnostics facility in Penzberg, Germany: Abbott Architect rHTLV-I/II, Ortho Avioq HTLV-I/II Microelisa system, Abbott Prism HTLV-I/HTLV-II, and DiaSorin Murex HTLV I+II. Fujirebio INNO-LIA HTLV-I/II Score was used as a confirmatory assay. The Elecsys HTLV-I/II, Abbott Architect rHTLV-I/II, and Abbott Prism HTLV-I/HTLV-II assays detected all HTLV-1/2-positive samples (sensitivity, 100%). Sensitivity for Ortho Avioq HTLV-I/II was 98.63%. The Elecsys HTLV-I/II assay had a specificity of 99.95% in blood donor samples, which was comparable to results for the other assays (range, 99.91 to 100%). In routine diagnostic samples, the specificity of the Elecsys HTLV-I/II assay was 99.83%, compared with 99.70% for Abbott Architect rHTLV-I/II. Specificity for the Elecsys HTLV-I/II assay in potentially cross-reactive samples was 100%, compared with 99.0% for Ortho Avioq HTLV-I/II and 99.2% for DiaSorin Murex HTLV I+II. The Elecsys HTLV-I/II assay has the sensitivity and specificity to support its use as a routine screening assay for detecting HTLV infection.

Highlights

  • Screening of blood for human T-cell lymphotropic virus type 1 and type 2 (HTLV-1 and -2, respectively) is important to diagnose and prevent infection and ensure the safety of blood supplies

  • One sample from the Barcelona center was excluded from the specificity calculations, having given the following results: Elecsys Human T-cell lymphotropic virus (HTLV)-I/II reactive, Abbott Architect rHTLV-I/II nonreactive, and Fujirebio INNO-LIA HTLV-I/II score reactive (“env gp21 I/II” [“1ϩ”], “env gp46 I/II,” “env gp46-I” [“ϩ/Ϫ”])

  • This study demonstrates that the Elecsys HTLV-I/II assay has the sensitivity (100%)

Read more

Summary

Introduction

Screening of blood for human T-cell lymphotropic virus type 1 and type 2 (HTLV-1 and -2, respectively) is important to diagnose and prevent infection and ensure the safety of blood supplies. The sensitivity and specificity of the Elecsys HTLV-I/II assay were determined using well-characterized HTLV-1/2-positive serum and plasma samples and routine diagnostic and blood donor samples expected to be HTLV negative, respectively. These results were compared with those for at least one of the following CE-marked assays at seven independent laboratories and the Roche Diagnostics facility in Penzberg, Germany: Abbott Architect rHTLV-I/II, Ortho Avioq HTLV-I/II Microelisa system, Abbott Prism HTLV-I/HTLV-II, and DiaSorin Murex HTLV IϩII. Specificity for the Elecsys HTLV-I/II assay in potentially cross-reactive samples was 100%, compared with 99.0% for Ortho Avioq HTLV-I/II and 99.2% for DiaSorin Murex HTLV IϩII. The Elecsys HTLV-I/II assay has the sensitivity and specificity to support its use as a routine screening assay for detecting HTLV infection

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)