Abstract

Fresh semen specimens from 46 patients and donors were evaluated for concentration, motility, velocity, and linearity using two different commercially available computerized sperm motion analyzer systems. Although no significant differences in measurement of concentration or motility were observed, significant differences in velocity and linearity were recorded. Fourteen cryopreserved/thawed samples were assessed with the same set-up parameters as fresh specimens. When discrepancies between manual and computer counts were noted, the authors changed the set-up parameters and evaluated 33 additional specimens. Again, no differences in concentration and motility, but significant differences in velocity and linearity were observed. Interlaboratory results must be correlated and standardization of set-up parameters of various analyzers is essential.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.